Adenosine-stress dynamic myocardial perfusion imaging using 128-slice dual-source CT: optimization of the CT protocol to reduce the radiation dose

Citations

WEB OF SCIENCE

34
Citations

SCOPUS

38

초록

The aim of this study was to compare the radiation dose and image quality of different adenosine-stress dynamic myocardial perfusion CT protocols using a 128-slice dual-source computed tomography (DSCT) scanner. We included 330 consecutive patients with suspected coronary artery disease. Protocols employed the following dynamic scan parameters: protocol I, a 30-s scan with a fixed tube current (FTC, n = 172); protocol II, a 30-s scan using an automatic tube current modulation (ATCM) technique (n = 108); protocol III, a 14-s scan using an ATCM (n = 50). To determine the scan interval for protocol III, we analyzed time-attenuation curves of 26 patients with myocardial perfusion who had been scanned using protocol I or II. The maximum attenuation difference between normal and abnormal myocardium occurred at 18.0 s to 30.3 s after initiation of contrast injection. Myocardial perfusion images of FTC and ATCM were of diagnostic image quality based on visual analysis. The mean radiation dose associated with protocols I, II, and III was 12.1 +/- A 1.6 mSv, 7.7 +/- A 2.5 mSv, and 3.8 +/- A 1.3 mSv, respectively (p < 0.01). Use of a dose-modulation technique and a 14-s scan duration for adenosine-stress CT enables significant dose reduction while maintaining diagnostic image quality.

키워드

Myocardial perfusion imaging; Dual-source computed tomography; Image quality; Radiation dose; CT protocol; TUBE CURRENT MODULATION; COMPUTED-TOMOGRAPHY; ANGIOGRAPHY
제목
Adenosine-stress dynamic myocardial perfusion imaging using 128-slice dual-source CT: optimization of the CT protocol to reduce the radiation dose
저자
Kim, Sung Mok; Kim, Yoo Na; Choe, Yeon Hyeon
DOI
10.1007/s10554-012-0138-x
발행일
2013-04
유형
Article
저널명
International Journal of Cardiovascular Imaging
권
29
호
4
페이지
875 ~ 884