The Proton Therapy Nozzles at Samsung Medical Center: A Monte Carlo Simulation Study using TOPAS

Citations

WEB OF SCIENCE

10
Citations

SCOPUS

10

초록

To expedite the commissioning process of the proton therapy system at Samsung Medical Center (SMC), we have developed a Monte Carlo simulation model of the proton therapy nozzles by using TOol for PArticle Simulation (TOPAS). At SMC proton therapy center, we have two gantry rooms with different types of nozzles: a multi-purpose nozzle and a dedicated scanning nozzle. Each nozzle has been modeled in detail following the geometry information provided by the manufacturer, Sumitomo Heavy Industries, Ltd. For this purpose, the novel features of TOPAS, such as the time feature or the ridge filter class, have been used, and the appropriate physics models for proton nozzle simulation have been defined. Dosimetric properties, like percent depth dose curve, spread-out Bragg peak (SOBP), and beam spot size, have been simulated and verified against measured beam data. Beyond the Monte Carlo nozzle modeling, we have developed an interface between TOPAS and the treatment planning system (TPS), RayStation. An exported radiotherapy (RT) plan from the TPS is interpreted by using an interface and is then translated into the TOPAS input text. The developed Monte Carlo nozzle model can be used to estimate the non-beam performance, such as the neutron background, of the nozzles. Furthermore, the nozzle model can be used to study the mechanical optimization of the design of the nozzle.

키워드

Monte Carlo simulation; Proton therapy; Passive scattering; Pencil-beam scanning; TOPAS
제목
The Proton Therapy Nozzles at Samsung Medical Center: A Monte Carlo Simulation Study using TOPAS
저자
Chung, Kwangzoo; Kim, Jinsung; Kim, Dae-Hyun; Ahn, Sunghwan; Han, Youngyih
DOI
10.3938/jkps.67.170
발행일
2015-07
유형
Article
저널명
Journal of the Korean Physical Society
권
67
호
1
페이지
170 ~ 174