Background: Early diagnosis of HIV infection reduces morbidity and mortality. Fourth-generation HIV detection assays are more sensitive because they can detect p24 antigen as well as anti-HIV antibodies. In this study, we evaluated the performance of a new fourth-generation ADVIA Centaur HIV antigen/antibody combo (CHIV) assay (Siemens Healthcare Diagnostics Inc., USA) for early detection of HIV infection and reduction of false positive rate. Methods: Four seroconversion panels were included. The third-generation ADVIA Centaur HIV 1/O/2 enhanced (EHIV) assay (Siemens Healthcare Diagnostics Inc., USA) and fourth-generation CHIV assay were used to test each panel for HIV infection. The presence of antigen was confirmed using HIV p24 antigen assay. To evaluate false-positivity and specificity, 54 HIV false-positive and HIV-negative serum samples from 100 hospitalized patients and 600 healthy subjects were included. Results: Compared to the EHIV assay, the CHIV assay had a shorter window for three of the seroconversion panels: a difference of 10 days and two bleeds in one panel, and 4 days and one bleed in the other two panels. Only 34 of the 54 (63%) samples known to yield false-positive results by EHIV assay had repeatedly yielded reactive results in the CHIV assay. One of the 600 healthy subjects had a false-positive result with the CHIV assay; thus, the specificity was 99.85% (699/700). CHIV accurately determined the reactive results for the HIV-confirmed serum samples from known HIV patients and Korea Food & Drug Administration (KFDA) panels. Conclusions: The new fourth-generation ADVIA Centaur HIV assay is a sensitive and specific assay that shortens the serological window period and allows early diagnosis of HIV infection.
키워드
- 제목
- Reduction of the HIV Seroconversion Window Period and False Positive Rate by Using ADVIA Centaur HIV Antigen/Antibody Combo Assay
- 저자
- Lee, Kyunghoon; Park, Hyung-Doo; Kang, Eun-Suk
- 발행일
- 2013-11
- 유형
- Article
- 권
- 33
- 호
- 6
- 페이지
- 420 ~ 425
- 언어
- ENG
- 출판사
- 대한진단검사의학회
- 발행국가
- 대한민국
- 분량
- 6 페이지
- ISSN
- E 2234-3814
P 2234-3806