A sensitive ocular perfusion apparatus measuring outflow facility

  • Kee, C; 
  • Hong, TM; 
  • Choi, K
Citations

WEB OF SCIENCE

10

초록

Purpose. To build an improved apparatus for measuring outflow facility by one- Methods. A computer-controlled and -monitored tower adjusts the height of a fluid reservoir by a stepper motor every 5 min, to keep intraocular pressure (IOP) constant at one or two desired pressure levels (e.g. 2.5 and 12.0 mm Hg above the spontaneous IOP). The fluid flow rate is calculated by the equation Q = (Pres - IOP)/R, where Pres is the pressure of the reservoir as calculated by the term ''rho gh,'' h being the height of the reservoir, measured by the rotation of stepper motor, g, the gravity constant, and rho, the fluid density. IOP is measured by the transducer, and R is the resistance of 180 mu m tubing. Outflow facility is calculated by Delta Q/Delta P. Delta Q is the difference between flow rate at the two pressure levels, and Delta P is the intraocular pressure difference. Perfusion was performed on live albino rat eyes by the two-level constant pressure technique, and on artificial steel eyes at a single constant pressure perfusion (20 nlm Hg). Results. Outflow facility of albino rat eyes and artificial steel eyes perfused for 40 and 30 min, respectively was 0.0507 +/- 0.0039 mu l/min/mm Hg and 0.3546 +/- 0.0034 mu l/min/mm Hg (mean +/- SD), respectively. Conclusions. Our perfusion apparatus is highly stable, sensitive, and capable of measuring subtle changes in outflow facility in vivo and in vitro.

키워드

artificial steel eye; constant pressure perfusion; ocular perfusion apparatus; outflow facility; rat eye; SYSTEM
제목
A sensitive ocular perfusion apparatus measuring outflow facility
저자
Kee, C; Hong, TM; Choi, K
DOI
10.1076/ceyr.16.12.1198.5023
발행일
1997-12
유형
Article
저널명
Current Eye Research
권
16
호
12
페이지
1198 ~ 1201