Feasibility Study of the Neutron Dose for Real-time Image-guided Proton Therapy: A Monte Carlo Study

Citations

WEB OF SCIENCE

3
Citations

SCOPUS

3

초록

Two full rotating gantries with different nozzles (multipurpose nozzle with MLC, scanning dedicated nozzle) for a conventional cyclotron system are installed and being commissioned for various proton treatment options at Samsung Medical Center in Korea. The purpose of this study is to use Monte Carlo simulation to investigate the neutron dose equivalent per therapeutic dose, H/D, for X-ray imaging equipment under various treatment conditions. At first, we investigated the H/D for various modifications of the beamline devices (scattering, scanning, multi-leaf collimator, aperture, compensator) at the isocenter and at 20, 40 and 60 cm distances from the isocenter, and we compared our results with those of other research groups. Next, we investigated the neutron dose at the X-ray equipment used for real-time imaging under various treatment conditions. Our investigation showed doses of 0.07 similar to 0.19 mSv/Gy at the X-ray imaging equipment, depending on the treatment option and interestingly, the 50% neutron dose reduction was observed due to multi-leaf collimator during proton scanning treatment with the multipurpose nozzle. In future studies, we plan to measure the neutron dose experimentally and to validate the simulation data for X-ray imaging equipment for use as an additional neutron dose reduction method.

키워드

Proton therapy; Monte-Carlo simulation; Neutron dose; MODULATED RADIATION-THERAPY; 2ND CANCERS; RADIOTHERAPY; EQUIVALENT; SYSTEM; RISK
제목
Feasibility Study of the Neutron Dose for Real-time Image-guided Proton Therapy: A Monte Carlo Study
저자
Kim, Jin Sung; Shin, Jung Suk; Kim, Daehyun; Shin, Eunhyuk; Chung, Kwangzoo; Cho, Sungkoo; Ahn, Sung Hwan; Ju, Sanggyu; Chung, Yoonsun; Jung, Sang Hoon; Han, Youngyih
DOI
10.3938/jkps.67.142
발행일
2015-07
유형
Article
저널명
Journal of the Korean Physical Society
권
67
호
1
페이지
142 ~ 146