Effective arrangement of separated transmit-only/receive-only RF coil for improvement of B<sub>1</sub> homogeneity at 7 Tesla

Citations

WEB OF SCIENCE

3
Citations

SCOPUS

3

초록

This article presents an effective arrangement with shifted transmit (Tx)-only and receive (Rx)-only (TORO) radiofrequency (RF) coils in a single-channel surface coil for improving the magnetic flux (B (1)) homogeneity in an ultra-high field (UHF) magnetic resonance imaging (MRI) scanner. The proposed new methodology for the coil arrangement using the shifted TORO RF coils was demonstrated for coils with 50-mm, 100-mm, and 150-mm-square surfaces and the results were compared to those for general Tx/Rx surface coils with the same dimensions. The computational analysis indicated that a homogeneous B-1 field was achieved when the Rx-only coil was shifted in the two-dimensional xy-plane away from the Tx-only coils. Because the proposed coil configuration provides a unique opportunity for increasing the B (1) homogeneity, this feature is likely to increase the feasibility via new coil arrangements of UHF surface design and fabrication.

키워드

MRI (Magnetic Resonance Imaging); RF (Radiofrequency) coil; TORO (Transmit-only/receive-only); Homogeneity; 7 Tesla; MACROSCOPIC SUSCEPTIBILITY; LINE ARRAY; MRI; PULSE; HEAD
제목
Effective arrangement of separated transmit-only/receive-only RF coil for improvement of B<sub>1</sub> homogeneity at 7 Tesla
저자
Im, Geun Ho; Seo, Jeong-Hoon; Kim, Kyoung-Nam; Heo, Phil; Chung, Julius Juhyun; Jang, Moon-Sun; Lee, Jung Hee; Kim, Jae-Hun; Kim, Sun I.
DOI
10.3938/jkps.65.616
발행일
2014-09
유형
Article
저널명
Journal of the Korean Physical Society
권
65
호
5
페이지
616 ~ 624