상세 보기
Performance Evaluation of the PowerChek SARS-CoV-2, Influenza A & B Multiplex Real-Time PCR Kit in Comparison with the BioFire Respiratory Panel
- Kim, Tae Yeul;
- Kim, Ji-Youn;
- Shim, Hyang Jin;
- Yun, Sun Ae;
- Jang, Ja-Hyun;
- ... Huh, Hee Jae;
- ... Kim, Jong-Won;
- 외 1명
WEB OF SCIENCE
9SCOPUS
10초록
Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) and influenza viruses may pose enormous challenges to our healthcare system. We evaluated the performance of the PowerChek SARS-CoV-2, Influenza A & B Multiplex Real-time PCR Kit (PowerChek; Kogene Biotech, Seoul, Korea) in comparison with the BioFire Respiratory Panels 2 and 2.1 (RP2 and RP2.1; bioMerieux, Marcy l'etoile, France), using 147 nasopharyngeal swabs. The limit of detection (LOD) of the PowerChek assay was determined using SARSCoV-2, influenza A, and B RNA standards. The LOD values of the PowerChek assay for SARS-CoV-2 and influenza A and B were 1.12, 1.24, and 0.61 copies/mu L, respectively. The positive and negative percent agreements of the PowerChek assay compared with RP2 and RP2.1 were 97.5% (39/40) and 100% (107/107) for SARS-CoV-2; 100% (39/39) and 100% (108/108) for influenza A; and 100% (35/35) and 100% (112/112) for influenza B, respectively. The performance of the PowerChek assay was comparable to that of RP2 and RP2.1 for detecting SARS-CoV-2 and influenza A and B, suggesting its use in diagnosing SARS-CoV-2 and influenza infections.
키워드
- 제목
- Performance Evaluation of the PowerChek SARS-CoV-2, Influenza A & B Multiplex Real-Time PCR Kit in Comparison with the BioFire Respiratory Panel
- 저자
- Kim, Tae Yeul; Kim, Ji-Youn; Shim, Hyang Jin; Yun, Sun Ae; Jang, Ja-Hyun; Huh, Hee Jae; Kim, Jong-Won; Lee, Nam Yong
- 발행일
- 2022-07
- 유형
- Article
- 권
- 42
- 호
- 4
- 페이지
- 473 ~ 477
- 언어
- ENG
- 출판사
- 대한진단검사의학회
- 발행국가
- 대한민국
- 분량
- 5 페이지
- ISSN
- E 2234-3814
P 2234-3806