Ultrasound-Guided Radiofrequency Ablation Using a New Electrode with an Electromagnetic Position Sensor for Hepatic Tumors Difficult to Place an Electrode: A Preliminary Clinical Study

Citations

WEB OF SCIENCE

3
Citations

SCOPUS

2

초록

Purpose To evaluate whether a new electrode embedded with an electromagnetic position sensor (EMPS) improves the technical feasibility of percutaneous radiofrequency ablation (RFA) in patients with hepatic tumors difficult to place an electrode under ultrasonography (US) guidance and to assess short-term therapeutic efficacy and safety. Materials and Methods This prospective study was approved by the institutional review board, and written informed consent was obtained from all patients. Between January 2015 and December 2016, 10 patients (7 men and 3 women; age range 52-75 years) with a single hepatic tumor (median 1.4 cm; range 1.1-1.8 cm) difficult to place an electrode under US guidance were enrolled. The technical feasibility of targeting and overlapping ablation during the RFA procedure was graded using a four-point scale and analyzed using the Wilcoxon signed rank test according to the use of EMPS. In addition, the rates of technical success, local tumor progression (LTP), and major complications were assessed. Results The use of the new RF electrode with EMPS significantly improved the technical feasibility of targeting and overlapping ablation (p = 0.002 and p = 0.003, respectively). After treatment, the technical success rate was 100%. LTP was not found in any patient during the follow-up period (median 8 months; range 4-22 months). No major procedure-related complications occurred. Conclusions The technical feasibility of percutaneous RFA improves with the use of this RF electrode embedded with an EMPS. Short-term therapeutic efficacy and safety after RFA using the electrode were promising in patients with hepatic tumors difficult to place an electrode under US guidance.

키워드

Liver; Ultrasonography; Radiofrequency ablation; Electromagnetic position sensor; Needle tracking; TERM THERAPEUTIC OUTCOMES; HEPATOCELLULAR-CARCINOMA; NEEDLE TRACKING; FUSION; GUIDANCE; BIOPSY; PROGRESSION; NAVIGATION; MARGIN; US
제목
Ultrasound-Guided Radiofrequency Ablation Using a New Electrode with an Electromagnetic Position Sensor for Hepatic Tumors Difficult to Place an Electrode: A Preliminary Clinical Study
저자
Kang, Tae Wook; Lee, Min Woo; Song, Kyoung Doo; Rhim, Hyunchul; Lim, Hyo Keun; Kang, Wonseok; Kim, Kyunga
DOI
10.1007/s00270-017-1751-9
발행일
2017-12
유형
Article
저널명
CardioVascular and Interventional Radiology
권
40
호
12
페이지
1891 ~ 1898