Cardiopulmonary Coupling Analysis Using Home Sleep Monitoring System Based on Air Mattress

Citations

WEB OF SCIENCE

2
Citations

SCOPUS

4

초록

This study investigates the feasibility of cardiopulmonary coupling (CPC) using home sleep monitoring system. We have designed a system to measure respiratory signals and normal-to-normal (NN) interval series in a non-contact based on air mattress. Then, CPC analysis was conducted using extracted respiratory signals and NN interval series, and six CPC parameters were extracted (VLFC, LFC, HFC, e-LFC, eLFCNB and e-LFCBB). To evaluate the proposed method, two statistical analyses were conducted between the CPC parameters extracted by the electrocardiogram-based conventional method and the air mattress-based proposed method for five patients with obstructive sleep apnea and hypopnea (OSAH). Wilcoxon's signed rank test on the CPC parameters of the two methods indicated no significant differences (p > 0.05) and Spearman's rank correlation analysis showed high positive correlations (r >= 0.7, p < 0.05) between the two methods. Therefore, the proposed method has the potential for performing CPC analysis using air mattress-based system.

키워드

Cardiopulmonary coupling (CPC); Respiratory signal; Normal-to-normal (NN) interval; Home sleep monitoring system; Airmattress
제목
Cardiopulmonary Coupling Analysis Using Home Sleep Monitoring System Based on Air Mattress
저자
Park, Jong-Uk; Urtnasan, Erdenebayar; Joo, Eun-Yeon; Lee, Kyoung-Joung
DOI
10.1007/s10916-017-0824-2
발행일
2017-11
유형
Article
저널명
Journal of Medical Systems
권
41
호
11