상세 보기
초록
PURPOSE. The present study was aimed to evaluate the influence of implant contact ratio and stiffness of implant-surrounding materials on the resonance frequency analysis (RFA) values. MATERIALS AND METHODS. Seventy resin blocks that had the different amounts (100, 50, 30, 15%) of resin-implant contact (RIC) were fabricated. Ten silicone putty blocks with 100% silicone-implant contact were also made. The implants with empty set5.0 mm x 13.0 mm were placed on eighty specimen blocks. The RFA value was measured on the transducer that was connected to each implant by Osstell Mentor. Kruskal-Wallis and Scheffe's tests (alpha=.05) were done for statistical analysis. RESULTS. The control resin group with 100% RIC had the highest RFA value of 83.9, which was significantly different only from the resin group with 15% RIC among the resin groups. The silicone putty group with 100% contact had the lowest RFA value of 36.6 and showed statistically significant differences from the resin groups. CONCLUSION. Within the limitations of this in vitro study, there was no significant difference in the RFA values among the resin groups with different RIC's except when the RIC difference was more than 85%. A significant increase in the RFA value was observed related to the increase in stiffness of material around implant.
키워드
- 제목
- <i>In vitro</i> evaluation of resonance frequency analysis values to different implant contact ratio and stiffness of surrounding material
- 저자
- Kwak, Mu-Seung; Kim, Seok-Gyu
- 발행일
- 2013-11
- 유형
- Article
- 권
- 5
- 호
- 4
- 페이지
- 428 ~ 433
- 언어
- ENG
- 출판사
- 대한치과보철학회
- 발행국가
- 대한민국
- 분량
- 6 페이지
- ISSN
- E 2005-7814
P 2005-7806