Detection of pulmonary edema in pigs: Storage phosphor versus amorphous selenium-based flat-panel-detector radiography

  • Kim, TS; 
  • Im, JG; 
  • Goo, JM; 
  • Lee, KH; 
  • Lee, YJ; 
  • 외 2명
Citations

WEB OF SCIENCE

9
Citations

SCOPUS

10

초록

PURPOSE: To compare diagnostic accuracy of soft-copy selenium-based digital radiographic images and soft-copy computed radiographic images obtained for detection of pulmonary edema in pigs. MATERIALS AND METHODS: Oleic acid was injected intraatrially into three pigs (weight, 20-25 kg) at doses of 0.04, 0.05, and 0.06 mL/kg to induce pulmonary edema. Thirty-seven sets of computed radiographic, digital radiographic, and thin-section computed tomographic (CT) scans were obtained every 20-30 minutes in three pigs over 4-6 hours. Images were masked for identity, randomly sorted, and displayed on a monitor. Four radiologists rated each image for presence of parenchymal opacities by using a dichotomous scoring system in two sessions. Presence of pulmonary edema was determined with thin-section CT and a severity scale. Intra- and interobserver variations were determined with the K statistic and the Z test and with the Cochran Q test and the McNemar test, respectively. True-positive, true-negative, false-positive, and false-negative rates were determined. McNemar test was used to determine statistical significance of differences in detection between computed and digital radiographic images. RESULTS: There was no significant intra- or interobserver variation, except for one pair of observers during the first interpretative session with computed radiographic images (P =.016, McNemar test). Overall sensitivity (92.1%) and diagnostic accuracy (90.2%) of digital radiography were significantly higher than those of computed radiography (79.6% and 83.4%, respectively) (P <.001 for sensitivity, P =.01 for diagnostic accuracy, McNemar test). In detection of minimal and mild pulmonary edema, sensitivity of digital radiography (84%) was significantly higher than that of computed radiography (58%) (P <.001). CONCLUSION: Soft-copy digital radiographic images are superior to soft-copy computed radiographic images obtained for detection of mild pulmonary edema in pigs. (C) RSNA, 2002.

키워드

animals; computed tomography (CT), high-resolution; experimental study; lung, edema; radiography, comparative studies; radiography, digital; radiography, selenium detector; radiography, storage phosphor; ACTIVE-MATRIX READOUT; DIGITAL CHEST RADIOGRAPHY; SCREEN-FILM RADIOGRAPHY; SOFT-COPY IMAGES; COMPUTED RADIOGRAPHY; RADIOLOGY; SYSTEM; CT; QUALITY; LESIONS
제목
Detection of pulmonary edema in pigs: Storage phosphor versus amorphous selenium-based flat-panel-detector radiography
저자
Kim, TS; Im, JG; Goo, JM; Lee, KH; Lee, YJ; Kim, SH; Kim, S
DOI
10.1148/radiol.2233011079
발행일
2002-06
유형
Article
저널명
Radiology
권
223
호
3
페이지
695 ~ 701