Physical properties of new collimator cone system for stereotactic radiation therapy developed in Samsung Medical Center

  • Kim, DY; 
  • Ahn, YC; 
  • Oh, DG; 
  • Choi, DR; 
  • Ju, SG; 
  • 외 2명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

2

초록

Purpose: A new collimator cone system has been developed at the Samsung Medical Center that overcomes some of the limitations of present commercially supplied collimator cones. The physical properties of the newly developed cone system are described in this report. Methods and Materials: The new cones have relatively larger aperture sizes (3.0-7.0 cm in diameter) and are 16 cm in length. Each new cone is fabricated with cerrobend alloy melted and poured into a stainless steel housing that is permanently fixed to a mounting plate. The mounting plate of the new cone is designed to insert into the wedge mount slot of the pantry head. The mechanical accuracy of the central axis of the cone pointing to the isocenter was tested using film. a steel ball positioned at the isocenter by the mechanical isocenter device. For the evaluation of beam flatness and penumbra, off-axis ratios at 5 cm depth were measured by film dosimetry using polystyrene phantom. Results: The average error of the mechanical isocenter was 0.27 mm (+/- 0.16 mm), The beam flatness was excellent in the central region of the beam, and the average penumbra width was 3.35 mm (+/- 0.25 mm), The new cone design has more clearance between the patient's head and the gantry, and can more easily be removed from the gantry head because it slides in and out of the wedge slot. This facilitates changing cone sizes during one treatment session, and makes the process of double exposure port films easier. Conclusions: A new collimator cone system for stereotactic radiation therapy has been developed. The mechanical accuracy and physical properties are satisfactory for clinical use, and the new design permits a wider range of clinical applications for stereotactic radiation therapy. (C) 2000 Elsevier Science Inc.

키워드

stereotactic radiation therapy; collimator cones
제목
Physical properties of new collimator cone system for stereotactic radiation therapy developed in Samsung Medical Center
저자
Kim, DY; Ahn, YC; Oh, DG; Choi, DR; Ju, SG; Yeo, IH; Huh, SJ
DOI
10.1016/S0360-3016(00)00611-8
발행일
2000-09-01
유형
Article; Proceedings Paper
저널명
International Journal of Radiation Oncology Biology Physics
권
48
호
2
페이지
541 ~ 544