Technical feasibility and biocompatibility of a newly designed separating stent-graft in the normal canine aorta

  • Weon, YC; 
  • Kang, SG; 
  • Chung, JW; 
  • Kim, YI; 
  • Park, JH; 
  • 외 1명
Citations

WEB OF SCIENCE

3
Citations

SCOPUS

5

초록

OBJECTIVE. The objectives of this study were to assess the performance of a newly designed separating stent-graft system with respect to the technical feasibility of transfemoral deployment, the maintenance of vessel patency, and stent deformity due to mechanical defects; and to evaluate its in vivo healing characteristics, including thrombus formation, and endothelial covering of the stent-graft when placed in the normal aorta of a canine model. CONCLUSION. The newly designed separating stent-graft allowed accurate deployment without migration. This animal study also provided an opportunity to examine the healing process associated with an ultrathin polyester fabric nitinol stent and showed predictable healing characteristics in the normal thoracic aorta in this canine model.

키워드

animal studies; aorta; endothelialization; nitinol stent; stent-graft; ENDOVASCULAR GRAFT; DISSECTED AORTA; PERCUTANEOUS MANAGEMENT; ISCHEMIC COMPLICATIONS; ANEURYSMS; PLACEMENT; REPAIR; NITINOL; SHEEP
제목
Technical feasibility and biocompatibility of a newly designed separating stent-graft in the normal canine aorta
저자
Weon, YC; Kang, SG; Chung, JW; Kim, YI; Park, JH; Lee, DY
DOI
10.2214/AJR.05.0683
발행일
2006-04
유형
Article
저널명
American Journal of Roentgenology
권
186
호
4
페이지
1148 ~ 1154