상세 보기
초록
An automated system was developed for the routine preparation of carbon-ii (C-11) labeled radiopharmaceuticals, which consisted of three major parts including [C-11]methylation of the precursor with [C-11] iodomethane ([C-11]CH3I), purification of the desired product and formulation of the final C-11 labeled radiopharmaceutical. The whole system included seven three-way slider valves, eleven solenoid valves, four pneumatic cylinders, a HPLC (High Performance Liquid Chromatography) system and a rotary evaporator. Using this system, we investigated the radiochemical synthesis of L-[methyl-C-11]methionine, which is the most widely used amino acid in tumor PET (Positron Emission Tomography) studies. The overall operation took 30-35 min including the production of [C-11]CH3I (10.5 min) and decay-corrected radiochemical yield was 25%. The automated system we described herein can be widely utilized for the preparation of many C-11 labeled radiopharmaceuticals and has been shown to be efficient, reliable and easy to operate.
키워드
- 제목
- Development of an automated system for the routine preparation of carbon-11 labeled radiopharmaceuticals
- 저자
- Oh, SJ; Choe, YS; Kim, YS; Choi, Y; Kim, SE; Lee, KH; Ha, HJ; Kim, BT
- 발행일
- 1998-09-20
- 유형
- Article
- 권
- 19
- 호
- 9
- 페이지
- 952 ~ 956
- 언어
- ENG
- 출판사
- 대한화학회
- 발행국가
- 대한민국
- 분량
- 5 페이지
- ISSN
- E 1229-5949
P 0253-2964