ABO Mistyping of <i>cis</i>-AB Blood Group by the Automated Microplate Technique

  • Chun, Sejong; 
  • Ryu, Mi Ra; 
  • Cha, Seung-Yeon; 
  • Seo, Ji-Young; 
  • Cho, Duck
Citations

WEB OF SCIENCE

8
Citations

SCOPUS

11

초록

Background: The cis-AB phenotype, although rare, is the relatively most frequent of ABO subgroups in Koreans. To prevent ABO mistyping of cis-AB samples, our hospital has applied a combination of the manual tile method with automated devices. Herein, we report cases of ABO mistyping detected by the combination testing system. Methods: Cases that showed discrepant results by automated devices and the manual tile method were evaluated. These samples were also tested by the standard tube method. The automated devices used in this study were a QWALYS-3 and Galileo NEO. Exons 6 and 7 of the ABO gene were sequenced. Results: 13 cases that had the cis-AB allele showed results suggestive of the cis-AB subgroup by manual methods, but were interpreted as AB by either automated device. This happened in 87.5% of these cases by QWALYS-3 and 70.0% by Galileo NEO. Genotyping results showed that 12 cases were ABO*cis-AB01/ABO*O01 or ABO*cis-AB01/ABO*O02, and one case was ABO*cis-AB01/ABO*A102. Conclusion: Cis-AB samples were mistyped as AB by the automated microplate technique in some cases. We suggest that the manual tile method can be a simple supplemental test for the detection of the cis-AB phenotype, especially in countries with relatively high cis-AB prevalence. (c) 2018 S. Karger GmbH, Freiburg

키워드

ABO; Mistyping; cis-AB; Automation; Microplate technique; ERYTHROCYTE-MAGNETIZED TECHNOLOGY; VARIANT ALLELE; SYSTEM; GENE; IDENTIFICATION; PHENOTYPE; MUTATION; KOREA
제목
ABO Mistyping of <i>cis</i>-AB Blood Group by the Automated Microplate Technique
저자
Chun, Sejong; Ryu, Mi Ra; Cha, Seung-Yeon; Seo, Ji-Young; Cho, Duck
DOI
10.1159/000475506
발행일
2018-01
유형
Article
저널명
Transfusion Medicine and Hemotherapy
권
45
호
1
페이지
5 ~ 10